Training VideosR300 Series Calibration
R300 Series Calibration
R300 Series Calibration

Training R320 and R323 using all firmware (K342, K344, K354, K356)

Click above images for enlarged view.
 
Key Highlights & Features
Product Description

    As some of the digital setup steps can affect calibration, the BUILD and OPTION settings must be configured before calibration is attempted.

    Performing a Digital Calibration with Test Weights

    The Zero setting (CAL:ZERO) specifies a gross zero point for the scale. The Span setting (CAL:SPAN) specifies a second point (preferably close to full scale) used to convert the A/D readings into weighing units (eg. kg). Select either of the Zero (CAL:ZERO) or Span (CAL:SPAN) calibration items. It is important that an initial Zero calibration is performed before any SPAN calibrations.

    See in Download Section: One Page Step by Step Calibration Guide and R300 Series Reference Manual

    Note: It should not be assumed that just because the instrument has successfully calibrated a scale, that the scale is correct for trade use. Always check the scale build against the approval specification.

     

    Smart Weighing
    Training R320 and R323 using all firmware (K342, K344, K354, K356)

     Direct mV/V Entry Calibration

    In applications where test weights are not easily available, it is possible to calibrate the instrument directly by entering the mV/V signal strength at Zero and full scale Span. The Direct Zero setting (CAL:DIR.ZER) specifies a gross zero point for the scale. The Direct Span setting (CAL:DIR.SPN) specifies the mV/V signal strength corresponding to an applied mass equal to the full scale reading. This calibration technique is not compatible with linearisation. Clearly the accuracy of this type of calibration is limited to the accuracy of the direct mV/V data.

    ...now that's smart weighing

    Product Downloads
    The following files are available for download for the R300 Series Calibration. Ensure you select the correct documentation for your unit's software version. Check back regularly for software and other updates.

    To download files, please click on a filename below. A save dialogue box should appear allowing you to specify your save location.
    If a save dialogue box does not appear, right click on the link and select save target as.

    Files

     Calibration One Page Step by Step Calibration One Page Step by Step - 31 Jul 2014 - 199KB (PDF) file - R300-681 
     R300 Series Reference Manual R300 Series Reference Manual - 31 Jul 2014 - 696KB (PDF) file - R300-677